Computers - HP 4062UX Installation Manual

Semiconductor process control system semiconductor parametric test system
Table of Contents

Advertisement

13. Set select code for series 300 computers or logical unit number for series 700

computers.

Use
setsysconfig
# /usr/pcs/etc/setsysconfig
If your system is con gured in a diskless cluster environment, you must perform this step
for each diskless cnode and server.
Set select code (logical unit number) as follows:
Interface bus for instrument
Interface bus for prober
14. Perform
hpib4062
Enter as follows:
# /usr/pcs/etc/hpib4062
If your system is con gured in a diskless cluster environment, you must perform this step
for each diskless cnode and server.
This command checks whether three cards exist: 7, 25, and 27 select code or logical unit
number.
15. (HP 4062F only) Customize /usr/pcs/etc/FLCONFIG le.
The FLCONFIG le includes the information on the device selectors (select code
HP-IB address) of the HP 8110As and on the connections between pulse generator units
and HF ports.
You must specify the correct connections between pulse generator (PG) units and HF ports
in this le. The default setting in this le assumes that ve HP 8110As (containing two PG
units each) are installed in the HP 4062F system, and the PG units are connected to HF
ports 1 to 10 as follows:
1080 ! HF1 PG11
1090 ! HF2 PG12
1100 ! HF3 PG21
1110 ! HF4 PG22
1120 ! HF5 PG31
1130 ! HF6 PG32
1140 ! HF7 PG41
1150 ! HF8 PG42
1160 ! HF9 PG51
1170 ! HF10 PG52
For example, if three HP 8110As are installed and each HP 8110A has two PG units,
the PG4
and PG5
x
1080 ! HF1 PG11
1090 ! HF2 PG12
1100 ! HF3 PG21
1110 ! HF4 PG22
1120 ! HF5 PG31
1130 ! HF6 PG32
5-12 Software Installation
command as follows:
)
27
)
25
command to con rm HP-IB cards.
lines as follows:
x
2
100 +
to
add