Conditioning time
%A %B %C %D FLOW PRESS [LC]
TIME
0.0
30
70
0
During the conditioning time, liquid is delivered under the
initial condition of execution step (program No. 2 in this
example).
Initial
%A %B %C %D FLOW PRESS [IN]
TIME
0.0
30
70
0
Liquid is delivered in the initial status (initial condition of
program No. 2 in this example).
Gradient
%A %B %C %D FLOW PRESS [GR]
TIME
0.0
40
60
0
When "start" is input in the initial status, gradient elution
(program No. 2 in this example) will start.
Conditioning time
TIME %A %B %C %D FLOW PRESS [LC]
0.0
10
90
0
After completion of the final step of link program, the
system returns to the first step (program No. 1 in this
example) and conditioning is carried out.
Delay time (initial)
%A %B %C %D FLOW PRESS [IN]
TIME
0.0
10
90
0
Liquid is delivered in the initial status (initial condition of
program No. 1 in this example).
When "start" is not input during the delay time, column
flushing in ⑩ will begin.
When "start" is input during the delay time, the system
returns to action in ④.
Column flushing
%A %B %C %D FLOW PRESS [FL]
TIME
0.0
10
90
0
3 - 37
0
1.000
5.5
0
1.000
5.5
0
1.000
5.5
0
1.000
5.5
0
1.000
5.5
0
1.000
5.5
C 2
C 2
C 2
C 1
1
C 1