(2)
Sequential program setting
(a) Setting method
As a method of setting a sequential program, the
setting procedure is shown below on the following
conditions; analysis time 5 minutes and injection of
120 samples of 10 μL each sequentially once.
STEP RACK VIAL
1
1
This display appears at the end of initialization after power-
on.
PROG
EDIT MODE (MODIFY=0, NEW=1)
0
Select MODIFY for editing the previous program or NEW for
creating a new program.
0 ENT or 1 ENT
STEP RACK
1
1
Input a rack code.
1 ENT
STEP RACK
1
1
Input the first vial number.
1 ENT
STEP RACK
1
1
Input the last vial number.
120 ENT
STEP RACK
1
1
Input an injection volume (μL).
10 ENT
(to be continued)
3 - 13
VOL(µL)
INJ
1
10.0
1/1
VIALS
VOL(µL)
1- 1
10.0
VIALS
VOL(µL)
1- 1
10.0
VIALS
VOL(µL)
1- 1
10.0
VIALS
VOL(µL)
1-200
10.0
TIME
[I]
0.0
[C]
INJ
TIME
1
5.0
INJ
TIME
1
5.0
INJ
TIME
1
5.0
INJ
TIME
1
5.0