Hg lamp
(e) Detach the connector between the Hg lamp and main
body, loosen two retaining screws and detach the Hg
lamp.
Fig. 5-13 Removal of Hg Lamp
(f)
Attach a new Hg lamp in place.
Fasten the lamp with the two retaining screws, then
connect the connector with the main body.
(g) Reattach the flow cell unit and light source cover in
their original positions.
(2)
Entry of Lamp Replacement Date
When the above procedure is finished, enter a lamp
replacement date (see 3.3.4). With this entry, the
cumulative on-time is set to "0" and the number of lightings
to "1".
5 - 19
Close-up view
Connector
Retaining screws