(c) Remove the Hg lamp.
Hg lamp
(d) Fit a new Hg lamp and insert it into the instrument.
(e) Attach the Hg lamp cover at the original position.
In succession, attach the lamp cover at the original
position.
(4)
Check the wavelength accuracy of Hg lamp, referring to
4.4.1.
(5)
Enter the lamp replacement date as follows.
GLP
(a) Press
GLP (LOGBOOK=1, KEY LOCK=2,
1
D2 CHECK=3, W CHECK=4, Hg CHECK=5)
(b) Press [numerical] key [1], and then the [ENT] key.
The following screen appears.
LAMP SELECT (D2=1, W=2,Hg=3)
1
(c) Press [numerical] key [3], and then the [ENT] key.
The following screen appears.
LOGBOOK (REPORT=1, RESET=2)
1
(d) Press [numerical] key [2], and then the [ENT] key.
The following screen appears.
Hg LAMP CHANGE (MM DD YYYY)
<1–12>
5 - 13
key, and the following screen will appear.
_9
15
2010
5.2.2