3.2.5 Other Settings

Input Numeral
1
PROCESSOR
2
ERROR OUT
3
LEAK SENSOR :
4
LAMP
5
GAIN
(a)
Setting the Full Scale of Processor Output
SET UP
UTLTY
1)
Press
2)
The utility setup screen appears.
Utility Setup Screen
UTILITY (OFFSET=1, TIME CONSTANT=2,
1
USE PROG=3, POLARITY=4, OTHERS=5)
Select OTHERS.
Press [numerical] key [5], and then the [ENT] key.
3)
The OTHERS setup screen appears.
OTHERS (PROCESSOR=1, ERROR OUT=2
1
LEAK SENSOR=3, LAMP=4, GAIN=5)
Select PROCESSOR.
Press [numerical] key [1], and then the [ENT] key.
4)
The PROCESSOR setup screen appears.
PROCESSOR (0.25AU/FS=1, 0.5AU/FS =2
3
3 - 15
Function
:
Set the full scale of processor
output.
:
Select the Enable or Disable
status of error signal output.
Select use or non-use of the leak
sensor.
:
Select ON/OFF status of the
D
/W lamp.
2
:
Select a gain so as to match the
flow cell to be used.
key.
1.0 AU/FS =3, 2.0AU/FS =4)