7-9-30

LOGIQ E9 Replacing New Fan Tray - Fails "GFI Fan Test"

LOGIQ E9 console models 5205000, 5205000-2, 5205000-3, 5205000-4 running Application Software
R2.x.x or earlier. The introduction of the new Fan Tray (P/N 5394406) fails system level diagnostic "GFI
Fan Test."
The "GFI Fan Test" diagnostic checks for basic fan movement at two set points. The diagnostic checks
fan rotation by applying two different control voltage Duty Cycles to the fan tray. The current test
implementation sets restrictive limits on the rotation RPM. These rotation limits need to be widened
(become less restrictive) which will allow for fan and supplier variation.
This Section will instruct how review the test results data and assess if the fan is bad. This diagnostic
failure does not affect normal operation, it will only occur when running diagnostics.
GFI Fan Test Results Detail
Figure 7-203 Upper Limit on New Fan Rack Assembly will be 2500
NOTE:
DO NOT replace any part unless the measured RPM falls outside of the new limit.
PROPRIETARY TO GE
. 7
EV
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 P
ROPRIETARY
S
M
ERVICE
ANUAL
7 - 345