LOGIQ E9 GTX HV (High Voltage) Failure on MRX Systems (cont'd)
NOTE:
Use "FatalHW" as a search string to find error in the error log. This will help to identify the error
explained in step 2 and narrow down to the board that is causing it.
GTX "numbers" use 1, 2, 3 and 4 as nomenclature in VPD, but 0 (zero), 1, 2 and 3 are used as
nomenclature in diagnostics. In this manual, 1, 2, 3 and 4 nomenclature is used to identify board, based
on location in the Card Rack. See: Figure 7-201 "Card Rack - GTX Board Location (MRX Configuration)"
on page 7-336.
2.) Error log: The logfile.txt will display an entry like the following:
HouseKeeping: GFI Hardware Exception: SubType 31:
empty-GTX3 HVFAULT. Severity : FatalHW
Thursday,Aug 11 15:58:41,2011; ; Info
setting RunBlock flag to false!
.
.
.
Thursday,Aug 11 15:58:41,2011; ; Debug
GTX_PS_CR = 0x0038
Thursday,Aug 11 15:58:41,2011; ; Debug
GTX3 @0x0412 TEMP_MONITOR = 0x0000
Thursday,Aug 11 15:58:41,2011; ; Debug
GTX3 @0x0413 GTX_ErrorStatus = 0x0000
Thursday,Aug 11 15:58:41,2011; ; Debug
GTX3 @0x0421 Status_Bus_REGA = 0x0000
Thursday,Aug 11 15:58:41,2011; ; Debug
GTX3 @0x0422 Status_Bus_REGB = 0x0000
Thursday,Aug 11 15:58:41,2011; ; Debug
still alive, errors=0x0138 [(THOR) - - (TS_OK) (TS_LEVEL_OK) GTX3 -
0x0000 - This specific issue is identified by status registers of all zero value.
As the log indicates, the GTX3 was generating the error.
How to Fix:
The board needs to be replaced with the Rev. D or later, to prevent this error from reccurring.
The Rev. D Boards have been terminated correctly to prevent floating voltages.
NOTE:
Label the GTX Boards prior to removal to reflect which slot each board was removed from, if
replacing one or two boards. Install the boards back in the original locations.
Label the Front Planes prior to removal to reflect which is the upper and which is the lower. Install the
Front Planes back in the original locations.
Service Stock of the GTX Boards have been purged and populated with Rev. D or later.
NOTE:
GFI console models do not have this problem because those signals are terminated correctly
in the Backplanes for GFI console models.
PROPRIETARY TO GE
. 7
EV
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 P
ROPRIETARY
-GFE Ch0 empty-GFE Ch1
; DFEManager(2980); HandleRunFlag:
; acqGfi(2952); WatchDog:WatchDog :
; acqGfi(2952); WatchDog:WatchDog :
; acqGfi(2952); WatchDog:WatchDog :
; acqGfi(2952); WatchDog:WatchDog :
; acqGfi(2952); WatchDog:WatchDog :
; acqGfi(2952); WatchDog:Watchdog
S
M
ERVICE
ANUAL
7 - 335