Incorrect System Time When Daylight Savings Time Changes

1.) From the touch panel, press Utility -> System.
2.) Select Date/Time.
3.) On the Date and Time Properties screen, select the Time Zone tab.
4.) Disable the Automatically adjust clock for daylight saving changes checkbox, select Apply.
5.) Select the Date & Time tab.
6.) Set the proper time for the system's location, and then select OK.
7.) Select Save.
7-9-10-6

Touch Panel Not Responding

Perform Touch Screen Verification Utility, check first before calibrating. Access the Service
interface (Click on the Wrench) and go to Diagnostics > Service Diagnostics > Service Diagnostics
> Touch Panel > TP Cal Verification.
Calibrate touch panel. Access the Service interface and go to Diagnostics > Service Diagnostics >
Service Diagnostics > Touch Panel > TP Calibration.
Reload software.
Replace Upper Panel Assembly.
7-9-10-7

Service Password Does Not Respond When System Set to Greek/Russian Language (SN 70392)

Although you may change the language settings/keyboard to respond to different languages, the
keyboard must to be set to English to enter the password into the Service Browser or Maintenance
Access window.
If you enter the service password into the Service Browser login or Maintenance login, and it does
not respond: select the link to return to the Login page, press ALT+SHIFT to toggle the keyboard
to English, and enter the service password again.
7-9-10-8
Z Movement fails
If the Top Console does not lower, but it can be raised, this indicates that one of the control switches
is stuck in "Up" position.
If you have to help the Top Console up when raising it, but the motor assistance works when
lowering the Top Console, it indicates that the gas spring inside the Z Mechanism is failing.
-
7-9-11
Probes
7-9-11-1

Probe Recognition

Check all the probes on all the ports. Remove all the probes, and then check each probe singly in
every port.
-
-
7-9-11-2
V Nav Inside Sensors Failure
In the event that the V Nav Inside sensors would fail, the probe would continue to be recognized as a
V Nav Inside probe and would continue to scan correctly. On entry into V Nav, the LOGIQ E9 would
report a failure to initialize or connect to the V Nav sensors.
V Nav inside probes could be used with external sensors. In the event of an initialization / connection
failure to the V Nav Inside sensor, the LOGIQ E9 will automatically switch to and try to initialize the
external sensors.
7 - 236
PROPRIETARY TO GE
. 7
EV
Replace the Z Mechanism.
If the problems persist with all the probes, replace the GRLY board.
If only one probe fails to be recognized, clean User Defs to check if the problem is caused by
corrupted presets. If the problem is not resolved, replace the probe.