D
5180483-100, R
IRECTION
Table 7-57
Description of Function
Ticking or Popping sound in
speakers after system has fully
booted (speaker noise during
boot up is common and not a
problem).
1. Single tick - about every
second that matches the blinking
of the DS2 LED on the BEP door.
2. Double ticks - about every two
seconds, no blinking LEDS.
7 - 234
PROPRIETARY TO GE
. 7
EV
Boot Up Sequence
1. The known causes of this noise is a blown fuse in the BEP EPS or the EPS
battery cable is not fully seated in the connector.
Check the fuse in the EPS on the BEP. Replace the fuse if it is blown. If the fuse
is not blown, re-seat the battery cable on the EPS circuit board. If the noise
does not go away, order a new battery pack assembly with cable.
2. Replace the I/O Board.