D
5180483-100, R
IRECTION
7-9-18-3
Key Log Files (cont'd)
7-9-18-5-7
POP-UP: System Voltage Fault – System will Shutdown
Condition: An internal voltage is above or below preset limits
Example: Voltages cannot be read:
Error
limits. UpperBound = -4.000
Error
limits. UpperBound = 0.000 [...]
Error
limits. UpperBound = 5.000 [...]
Error
limits. UpperBound = 112.000 [...]
Error ; acqHouseKeeping(3400); GRLY_12V = -2147483.750 outside tolerance limits.
UpperBound = 12.600 [...]
Error ; acqHouseKeeping(3400); GRLY_VCC = -2147483.750 outside tolerance limits.
UpperBound = 5.250 [...]
Error ; acqHouseKeeping(3400); GRLY_3V3 = -2147483.750 outside tolerance limits.
UpperBound = 3.465 [...]
Error
limits. UpperBound = 3.900 [...]
Error
limits. UpperBound = 3.900 [...]
Error
Error
Info
monitoring event (strObjId = SYS_VOLTAGE_UNREADABLE, strData = SystemVoltage:
GRLY_LVDC = -2147483.750
UpperLimit = 3.900 LowerLimit = -1.000)
Error
the user to shutdown the system
In reality – voltages could not be read – monitor for further failures and take action if repeatable.
7 - 268
PROPRIETARY TO GE
. 7
EV
; acqHouseKeeping(3400);
; acqHouseKeeping(3400);
; acqHouseKeeping(3400);
; acqHouseKeeping(3400);
; acqHouseKeeping(3400);
; acqHouseKeeping(3400);
; acqHouseKeeping(1928); voltageDataOk = true
; acqHouseKeeping(1928); voltageErrorCount = 0
; acqHouseKeeping(1928); CHouseKeeping::GenerateMonitoringEvent:
; acqHouseKeeping(1928); Fatal error detected by HouseKeeping, prompting
Section 7-9 - Troubleshooting Tips
LOGIQ E9 P
GRLY_AVEE = -2147483.750 outside tolerance
[...]
GRLY_DLP_AUX = -2147483.750 outside tolerance
GRLY_PMX_VNN = -2147483.750 outside tolerance
GRLY_PMX_VPP = -2147483.750 outside tolerance
GRLY_LVDC = -2147483.750 outside tolerance
GRLY_LVDC = -2147483.750 outside tolerance
Volt is outside tolerance. Voltage is unreadable!