Network Troubleshooting (cont'd)
7-8-10-1
No Contact With Server, But Connection to Network is OK
Check Network setup on the scanner. See:
7-8-11

Volume Navigation (V Nav) Troubleshooting

If the distortion is not corrected by following the calibration instructions, you can check the functionality
of the sensors without the application running by using the following utility and observing the behavior
of the sensors:
C:HARDWARE_DRIVERS\drivebay\Cubes.exe
Section 7-9

Troubleshooting Tips

Each entry in the table will hyper-link to the issue and resolution.
Table 7-45
4D
Backend Isolation from Front End
(Simulator Mode)
Back End Processor
Also see:
Processor (BEP) Troubleshooting"
on page
PROPRIETARY TO GE
. 7
EV
Troubleshooting Tips Table
Sub-System
7-9-9 "Back End
7-223.
Chapter 7 Diagnostics/Troubleshooting
7-5-21-5 "Network Status" on page 7-139
See:
7-9-14 "4D Option Troubleshooting" on page 7-238
To isolate the Front End from the Backend and boot up into Simulator Mode,
you must:
• Power Down the System,
• GFI Configuration - Disconnect the PCIe cable on both the GFI Board and
BEP.
• MRX Configuration - Disconnect the PCIe cable to the Motherboard (J3).
• Disconnect the USB cable to the Patient I/O Module if present. This step
may be necessary since the Patient I/O Board is a self-powered USB device
and back powers the USB data lines causing boot up issues while the
system is disconnected from the front end.
In simulator mode, the scan screen may display pulsatile circles or no scan
at all, but will allow other application functions to run normally.
See:
7-9-8-1 "System Halt Errors - Lock ups or Intermittent Problems" on
(continued)
page 7-221
See:
7-9-8-2 "CD/DVD Drive Failures" on page 7-221
See:
7-9-8-3 "Image CD/DVD not read" on page 7-221
See:
7-9-8-4 "BEP Arrives Without Video" on page 7-221
See:
7-9-8-5 "Simulator Mode" on page 7-222
See:
7-9-8-6 "Lockups or Blue screens" on page 7-222
See:
7-9-8-7 "Files corrupt on boot" on page 7-222
See:
7-9-8-8 "Here's a way to power on the BEP" on page 7-222
See:
7-9-8-9 "Hard Drive Failure" on page 7-222