7. Checks Before and After Use

In the interests of safety, it is the user's responsibility to carry out the following checks
before and after using the system.
7.1
Checks Before Turning ON the Power
Before turning ON the power, perform the following checks.
No.
1
The temperature, humidity, and atmospheric pressure should meet the
conditions of use.
2
There should be no condensation.
3
There should be no deformation, damage, or stains to the system and
peripheral units.
* If any stains are present, perform cleaning referring to subsection 21.1.1
"Cleaning the system".
4
There should not backlash or loose screws in the casters, monitor, panel,
etc.
5
Caster locks should function correctly.
6
There should be no damage to cables (power cable, ECG cable etc.) and
no looseness in the connectors.
7
There should be no damage or stains to the transducer, transducer cables,
connector, etc.
* If any stains are present, perform cleaning, disinfection, or sterilization
referring to the operation manual provided with the transducer.
8
No clips etc. are placed on the main panel.
9
There should be no obstacles near the movable sections and air filter of
the system.
Check item
No. 2B730-973EN*D
7-1
Check column