CAUTION:
(2) Do not turn the system power OFF while a recording media is being
accessed. If the system power is turned OFF during access to the hard
disk, floppy disk, or magneto-optical disk, the disk may be damaged and
recorded data may be lost.
(3) If the power cannot be turned OFF by the normal procedure, turn OFF the
main switch on the power supply panel at the rear of the system.
These methods should not be used under normal conditions. They may
damage the system.
7. For some transducers, the specified operating and storage conditions are
stricter than those for the diagnostic ultrasound system. Refer to the operation
manual provided with the transducer and check the specified environmental
conditions for operation and storage.
8. Turn OFF the system subswitch or stop transmission by FREEZE switch before
connecting or disconnecting a transducer. If a transducer is connected or
disconnected with an image displayed, the system and/or the transducer may
malfunction.
9. The user can record ID registration and user registrations (with a function to
permit user-specific items such as the hospital name to be registered). Be sure
to back up such information on external storage media. Data stored in the
system may be lost due to improper operation or an accident.
10. Observe the following precautions before moving the system to prevent system
malfunctions.
(1) Turn OFF the sub switch and the main switch on the power panel on the
rear of the system.
(2) Store the transducer in the transducer holder and hang the cable on the
transducer cable hanger.
(3) When moving the system, use the handle. Do not push weak sections
such as the panel.
11. Move the system by the handle. If the user pushes or pulls other sections, the
system may be damaged.
12. When moving the LCD monitor, be careful to avoid hitting the monitor against
the cable arm. The components may be damaged.
13. Do not press or use force on the main panel. It may damage the system.
14. Do not boot the system while the air filters are removed. The system may
malfunction.
15. If this system is used in a small room, it may cause the room temperature to
rise. Proper ventilation must be provided.
16. Do not connect or disconnect a printer from the Deskjet series when the system
power is ON. The printer may become defective.
17. The air filter at the left of the system must be cleaned once every 6 months. If
the air filter is clogged, the internal temperature will rise, shortening the service
life of the system. For inspection and cleaning by service personnel, contact
your TOSHIBA representative.
No. 2B730-973EN*D
S-10