9. Entering the Patient ID

CAUTION:
Observe the following precautions to avoid erroneous
diagnosis.
1. Before examining a new patient, press
patient information and data recorded in the image memory
for the previous patient. Otherwise, the new data may be
confused with the data of the previous patient.
2. If the patient information is modified by pressing
patient information for the data already recorded in the
image memory is not modified (the data recorded in the
image memory is not deleted). Thus, if the data recorded in
the image memory is that for the previous patient, images
for the different patients will have the same patient
information, and this may cause confusion.
It a patient ID is to be changed, press
patient information and data recorded in the image memory
for the previous patient. Otherwise, the data belonging to
the current patient may be confused with the data for the
previous patient.
3. Confirm that the patient to be examined is the same as the
patient ID. There is a possibility of recording images with
incorrect patient IDs.
4. Enter the patient's height and weight correctly in the patient
ID registration screen. If the entered height and weight are
not correct, the BSA will be calculated incorrectly.
5. Enter the patient information including LMP and IVF
correctly in the patient ID registration screen. Otherwise, the
fetal growth data will be calculated incorrectly.