8. Checks Before and After Use

In the interests of safety, it is the user's responsibility to carry out the following checks
before and after using the system.
8.1

Checks Before Turning ON the Power

Perform the following checks before the system power is turned ON and after the
system power is turned OFF.
No.
1
The temperature, humidity, and atmospheric pressure should meet the
conditions of use.
2
There should be no condensation.
3
There should be no deformation, damage, or stains to the system and
peripheral units.
* If any stains are present, perform cleaning referring to subsection 20.3.1
"Cleaning the system".
4
There should not backlash or loose screws in the casters, monitor, panel,
etc.
5
Caster locks should function correctly.
6
There should be no damage to the reference signal cable and no
looseness of the connector.
7
There should be no abnormalities such as damage, wear, or bent pins on
the power cable or plug.
8
There should be no abnormalities or stains of the transducer surface, cable
sheath, or connector, such as peeling, cracks, protruding parts, or loss of
firmness of the acoustic lens.
* If any stains are present, perform cleaning, disinfection, or sterilization
referring to the operation manual provided with the transducer.
9
No clips etc. are placed on the main panel or keyboard.
10
There should be no obstacles near the movable sections and air filter of
the system.
Check item
No. 2B771-004EN*M
8-1
Check column