3.
Safety Precautions
Please observe the following precautions to ensure patient and operator safety when using this
system.
DANGER:
Do not use flammable gases such as anesthetics or flammable liquids
such as ethanol near this product. Doing so may cause an explosion.
This product is not intended for use with oxygen or in an oxygen
enriched atmosphere.
WARNING:
1. Be sure to connect the power plug to a two-pin (with dual earth
2. When in the patient environment, the operator must not touch
3. Information concerning the terminal to which the potential
4. Information concerning the functional earth terminal
5. Connect the earth conductor only before turning ON the system.
6. Do not connect this system to outlets with the same circuit
terminal) outlet meeting the ratings indicated on the rating
nameplate. If this type of outlet is not available, contact your
TOSHIBA representative.
any exposed connectors. In addition, if the system covers have
been removed for some reason, the operator must be extremely
careful not to touch any part where the voltage exceeds 25 VAC
or 60 VDC and the patient at the same time.
equalization conductor is to be connected (
To use another medical device in combination with this system,
an equipotential wire for connecting to an equipotential bus must
be supplied. For more information, contact your TOSHIBA
representative.
* Be sure to connect the potential-equalization lead wire before
inserting the equipment power plug into the receptacle. Also,
be sure to remove the equipment power plug from the
receptacle before disconnecting the wire to avoid electrical
shock.
(
):
Use of the functional earth terminal as protective earth is not
allowed. The system should be connected according to local
requirements and should not conflict with IEC60601-1
requirements.
Disconnect the grounding cable only after turning OFF the
system. Otherwise, electric shock may result.
breakers and fuses that control current to devices such as life-
support systems. If this system malfunctions and generates an
overcurrent, or when there is an instantaneous current at power
ON, the circuit breakers and fuses of the building's supply circuit
may be tripped.
No. 2B730-973EN*D
S-2
):