Toshiba famio XG SSA-530A Operation Manual page 16

Diagnostic ultrasound system
Table of Contents

Advertisement

CAUTION:
3. Precautions concerning installation and movement of the system
(1) Be sure to install the system on a level floor and lock the
(2) Do not push the system from the side. If the system is pushed
(3) When the system is moved over a sloped surface, it must be
(4) Do not sit on the system.
(5) Do not place any objects on top of the monitor. They may fall,
(6) Confirm that the peripheral units are secured before moving the
(7) If the ECG cable is connected, disconnect it before moving the
(8) When the system is moved over a step, be careful not to allow
4. To prevent electric shock, do not connect the peripheral units (video
printer, VCR, etc.) to an external outlet (except for Deskjet series
printers). Peripheral units should be connected to the service outlet
of the system. For the connection procedures, contact your
TOSHIBA representative.
In addition, be sure to use the cable supplied with the system to
connect a Deskjet series printer. If a different cable is used, there is
a risk of electric shock.
5. Before cleaning the system, be sure to disconnect the power cable
from the outlet. If the system is defective, there is a risk of electric
shock.
6. Do not allow fluids such as water to contact the system or
peripheral devices. Electric shock may result.
7. Terminals and switches on the system rear panel
(1) Only equipment that conforms to the safety standards must be
casters. If this is not done, the system may move, injuring the
patient.
from the side, it may fall down and cause injury.
moved slowly by two persons. Otherwise, the system may slide
unexpectedly and cause a serious injury.
The system may move, causing you to lose your balance and
fall.
causing injury.
system. Otherwise, the peripheral units may fall and cause
injury.
system. Otherwise, the cable may get caught in the casters or
someone may trip over the cable and injury may result.
the system to fall. When holding the system at the bottom to
help move it over a step, take special care to prevent hand
injuries.
connected to the ETHERNET terminal on the rear panel.
Otherwise, smoke or an electric shock may result.
No. 2B730-973EN*D