GE B20 User Manual page 84

Patient monitor
Hide thumbs Also See for B20:
Table of Contents

Advertisement

Respiration equipment to patient connection
Use the same setup as in the ECG measurement, see "ECG" section.
NOTE: Impedance respiration measurement with neonates is not
available in specific regions, including USA, Argentina, Guam, Puerto
Rico, Saint Croix, Saint Thomas, and Canada.
Turning on the respiration measurement
Select respiration to a waveform or digit field, otherwise respiration
data is not included in trends and no alarms are activated.
Others
1. Press the
key.
2. Select Resp Setup - Measurement - ON.
Selecting the respiration waveform size
Others
1. Press the
key and select Resp Setup.
2. Select Size.
The greater the value, the larger the waveform size.
80
Selecting the respiration rate source
Others
1. Press the
key
2. Select Resp Setup - Resp Rate Source
3. Select the appropriate option:
AUTO
CO2
Imped.
NOTE: The AUTO selection lets the monitor choose the
respiration rate source. If CO
used. If not, then impedance respiration is chosen as the
respiration rate source.
Adjusting the respiration detection limit
Normally, the AUTO detection limit is recommended. However, if the
respiration is weak or affected by artifacts, they may not be included
in the respiration rate. To ensure the correct respiration rate, adjust
detection limits:
Others
1. Press the
key.
2. Select Resp Setup - Detection Limit.
3. Adjust the limits.
Turning off the respiration measurement
Others
1. Press the
key
2. Select Resp Setup - Measurement - OFF.
is monitored, it is chosen to be
2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

B40