Select inflation limits
−
NIBP: You are using a hose without an automatic
identification. Select appropriate inflation limits.
NOTE: AUTO option is not available for these hoses.
•
Snapshot created
−
A snapshot is created manually or on alarms.
•
Snapshot memory full. Oldest snapshot erased.
−
The oldest snapshot lost.
•
SpO2 probe off
−
Check connection between sensor and patient.
−
Replace the sensor.
•
SpO2 faulty probe
−
Replace the sensor.
•
SpO2 module error
−
Contact authorized service personnel.
•
Tachy
−
Check the patient status.
•
Temp sensor error
−
Contact authorized service personnel.
•
T1 temperature error/T2 temperature error
−
Check the probe whether is set to 400.
−
Contact authorized service personnel.
•
V Fib
−
Check the patient status.
128
•
V Tach
−
Check the patient status.
•
Weak pulsation
−
Check the patient status.
−
Check the NIBP cuff position and attachment.
−
Check that the cuff is not damaged.
−
Repeat the measurement and check the patient status.
•
Zero ICP separately
−
In IBP menu to zero ICP labelled channel separately.
•
xxx high/low (xxx = measurement parameter)
−
Check the patient status.
−
Adjust the alarm limits.
•
xxx measurement removed
−
Acquisition module has been removed. Connect the module
if you want to restart the measurement.
−
Check the patient status.