GE B20 User Manual page 147

Patient monitor
Hide thumbs Also See for B20:
Table of Contents

Advertisement

*
This information may be useful to clinicians, such as those performing
photodynamic therapy.
Invasive blood pressure
Measurement range
Measurement accuracy
Frequency response
Transducer sensitivity
Pulse rate
Range
Accuracy
Display resolution
Zero adjustment range
-40 to 320 mmHg (-5.3 to 42.7 kPa)
±5% or ±2 mmHg, whichever is
greater
4 to 22 Hz
5 µV/V/mmHg
30 to 250 bpm
±5% or ±5 bpm, whichever is greater
1 bpm
± 150 mmHg
NIBP
Measurement technique
Supported modes
Measurement time
Measurement ranges
Systolic
MAP
Diastolic
Accuracy
Default initial inflation
pressure
Over pressure allowed by
independent safety
controller
1. Blood pressure measurements determined with this device are
equivalent to those obtained by an intra-arterial blood pressure
measurement device, within the limits prescribed by the American National
Standard, Manual, electronic, or automated sphygmomanometers
143
Oscillometric with step deflation
Manual, automatic and stat
Adult/Pediatric inflate duration time
less than 120 s
Neonate cycle time less than 85 s
Adult/Pediatric: 30 to 290 mmHg
Neonate: 30 to 140 mmHg
Adult/Pediatric: 20 to 260 mmHg
Neonate: 20 to 125 mmHg
Adult/Pediatric: 10 to 220 mmHg
Neonate: 10 to 110 mmHg
According to AAMI SP10-2002 4.4.5.2
B, accuracy of NIBP parameter was
validated against the intra-arterial
method
1
.
Adult/Pediatric: 135 ± 15 mmHg
Neonate: 100 ± 15 mmHg
Adult/Pediatric: 300 ± 6 to 330 mmHg
Neonate: 150 ± 3 to 165 mmHg