Batteries
Warnings
EXPLOSION OR FIRE-Using non-recommended batteries could
result in injury/burns to the patients or users. Only use batteries
recommended or manufactured by GE.
EXPLOSION HAZARD - not suitable for use in the presence of a
flammable anaesthetic mixture with air or with oxygen or nitrous
oxide
EXPLOSION HAZARD - Do not incinerate the battery or store at
high temperatures
Do not disassemble, open, or shred the battery; do not short
circuit the battery pack; charge only with internal charger
The monitor supports up to two lithium-ion batteries. Each battery
can be charged separately. During charging, screen symbols and
monitor frame LEDs indicate the current charging level and any
possible failures. You can check the battery status through
Setup
- Battery Setup. For additional information please refer to the
table following this statement.
The battery should be used if the integrity of the PROTECTIVE EARTH
CONDUCTOR in the installation is in doubt.
To configure to have the battery charge status visible at all time,
follow the following steps:
Fields - Battery.
NOTE: When the monitor is battery powered, the green battery LED
is on. When the monitor is connected to the mains power, the green
mains LED is on. See also sections
"Messages."
NOTE: If batteries are installed and the monitor is disconnected from
the mains power, it will automatically switch to battery power
without interruption of operation.
25
Monitor Setup
- Screen Setup - Digit
"Conditioning the
batteries" and
Monitor