No printer selected
Select a printer, see the
Patient discharged
Discharge patient.
Patient admitted
Admit patient.
P1 over range/P2 over range
Check the patient status.
P1 under range/P2 under range
Check the patient status.
Check the cables.
Rezero the transducer.
Change the module.
If the problem persists, contact authorized service personnel.
Printer error
Printer is not working properly. Check that the network
printer is operational.
Printing...
Printing on network printer has been started. Please wait
until the printing is finished.
Printing ready
Monitor has completed sending printing data to the printer.
Please wait until the printing is finished.
Recorder: out of paper
Add paper.
"Printing and
recording" chapter.
Recorder: cover open
Close the recorder cover.
Recorder: input voltage high, Recorder: input voltage low
Contact authorized service personnel.
Recorder: thermal array overheat
Contact authorized service personnel.
Recorder module removed
Reconnect the recorder module if you need a recorder.
Recorder: system error 29
The message rarely appear when the record information too
large, the current recording will stop also. Please press the
Recorder Start/Stop
Recorder: system error X
Contact authorized service personnel.
Replace Battery A, Replace Battery B
Replace the battery as soon as possible, see
batteries."
Replace D-Fend
Gases: Replace the water trap.
Restart needed
The monitor should be restarted.
Saving to network
Saving trend and information to Network
Sample line blocked
Gases: Change the sampling line and water trap.
127
key again.
"Replacing the