Factory default descriptions
The channels have the following factory default descriptions:
LABEL
IBP1, Art, ABP IBP2, CVP
Scale
mmHg/kPa
200/30
Color
Red
Alarm source
Sys
Digit format
S/D
Filter (Hz)
22
Response
Normal
Combinining pressures
All invasive pressure waveforms can be displayed together so that
they use an area of two normal waveforms, or so that all are
combined in the same field with the same zero line.
Monitor Setup
1. Press the
2. Select Screen Setup.
3. Select Waveform Fields.
To combine all pressure waveforms in one field, select
Combine Pressures to YES.
RAP, LAP
20/4
20/4
Blue
White
Mean
Off
Mean
Mean
9
9
Normal
Normal
key.
98
ICP
PA
20/4
60/8
White
Yellow
Off
Off
CPP
S/D
9
9
Normal
Normal
Determining pressure values visually
By moving the horizontal cursor across the pressure waveform, you
can get accurate pressure values at selected points. This may be
useful, for example, if the patient's breathing pattern is irregular. The
cursor is not available for pressures shown with a combined scale.
IBP
1. Press the
key.
2. Select IBP1 Setup - IBP1 Cursor and IBP2 Setup - IBP2 Cursor
3. Move the cursor up or down by turning the Trim Knob. Every time
the cursor is moved, the time (hours and minutes) and pressure
values appear on the screen. In this way you can keep track of
the changes made.
4. You can remove the cursor by selecting Remove Cursor. Note
that if the cursor is not removed, it remains visible on the Normal
Screen.
RVP
UAC
60/8
100/14
White
Red
Off
Sys, Dia,
Mean
S/D
S/D
9
14
Normal
Normal
UVC
10/2
White
Mean
Mean
14
Normal