OCRG
The monitor supports 8 minutes OCRG (oxycardiorespirogram)
function in the NEONATAL mode. The OCRG subsystem provides
services to view and review specific high resolution trends, high
resolution beat-to-beat HR, high resolution beat-to-beat SpO
compressed respiration waveform - simultaneously in the same
view. Two types of views are provided: OCRG Snapshot view and
OCRG Realtime view.
View OCRG snapshot
The OCRG snapshot is created when an OCRG event is triggered.
Refer to
"Setup OCRG"
for OCRG event setup. The monitor can store
up to 70 OCRG snapshots. OCRG snapshot includes the trend data 6
minutes before and 2 minutes after the OCRG event. If multiple
OCRG events are triggered within 2 minutes, they will be detected as
one OCRG event. During this time, only the first OCRG event will
trigger OCRG Snapshot creation.
To display the OCRG snapshot record.
1. In NEONATAL mode, press the
2. Select OCRG Snapshots.
You can select Scroll Snapshots and scroll left and right with the
Trim Knob to see the earlier OCRG snapshots. The trigger time and
condition also showed on the left menu area.
You can select Remove Menu to hide the left side menu. Press the
Trim Knob to show the menu.
and
2
Pt.Data & Trends
key.
54
View realtime OCRG
To display the realtime OCRG.
1. In NEONATAL mode, Press the
2. Select OCRG Realtime.
You can select Remove Menu to hide the left side menu. Press the
Trim Knob
to show the menu.
Setup OCRG
The OCRG event is that any one of HR, SpO2, Apnea is out of OCRG
triggering limits range. You can setup these limits.
Adjust OCRG alarm limits
1. In NEONATAL mode, Press the
2. Select OCRG Setup.
3. Select Adjust Limits.
Adjust short apnea time
1. In NEONATAL mode, Press the
2. Select OCRG Setup.
3. Select Short Apnea.
NOTE: The default short apnea time is 20 seconds.
Pt.Data & Trends
key.
Pt.Data & Trends
key.
Pt.Data & Trends
key.