Spray Arm Monitoring/Wash Pressure - Miele PG 8581 Operating Instructions Manual

Dental washer-disinfector
Table of Contents

Advertisement

Frequently asked questions

Spray arm monitoring/wash pressure

Problem
Spray arm monitoring - Upper
spray arm: Spray arm blocked
or excessive foaming
or
Spray arm monitoring - Lower
spray arm: Spray arm blocked
or excessive foaming
or
Spray arm monitoring - mobile
unit spray arm 1 - : spray arm
blocked or excessive foaming
Spray pressure exceeds tol-
erance
Spray pressure fluctuating too
much
122
Possible cause and solution
The set spin speed has not been reached.
– Wash items are obstructing the machine or basket spray
arms.
Arrange the wash items so that the spray arms can turn
easily and start the program again.
– The relevant spray arm is blocked.
Clean the spray arm.
Check whether the filters in the wash cabinet are clean
and correctly inserted.
Start the program again.
– Wash pressure is too low due to a heavy build-up of
foam.
Follow the instructions regarding foam build-up in
"Chemical processes and technology".
Rinse
Start the
program to clean the wash cabinet.
Then reprocess the wash items again.
The wash pressure differs from the reference value.
Possible causes of fluctuations in the wash pressure in-
clude:
- defective water connections,
- open adapters,
- foam build-up.
Identify and resolve the cause of this.
The program is not interrupted. Nevertheless, you must
reprocess the load.
A program was interrupted because of severe fluctuations
in the wash pressure.
Possible causes of fluctuations in the wash pressure in-
clude:
- defective water connections,
- open adapters,
- foam build-up.
Identify and resolve the cause of this.
Reprocess the load again.