Maintenance Measures; Maintenance - Miele PG 8581 Operating Instructions Manual

Dental washer-disinfector
Table of Contents

Advertisement

Maintenance measures

100

Maintenance

Periodic checks must be carried out by Miele Service after 1,000 op-
erating hours or every 24 months at the latest.
Maintenance covers the following:
- Electrical safety according to national regulations
- Door mechanism and door seal
- Any screw connections and connectors in the wash cabinet
- Water inlet and drainage
- Internal and external dispensing systems
- Spray arms
- Filter combination
- Sump including drain pump and non-return valve
- All mobile units, baskets, modules and inserts
- Steam condenser
- Wash pressure sensor
If there is a communication module:
- Connected printer
- network connection
External documentation software and computer networks are not
tested by Miele Service.
The following operational tests will be carried out within the frame-
work of the maintenance:
- A program will be run as a test run
- Seals will be tested for water tightness
- All relevant measuring systems will be safety tested, including fault
displays
- Safety features