Problem
Fault 555
Fault 559
Fault 578
Possible cause and solution
Too much water has accumulated in the steam condenser.
Wash water may have been diverted behind the protective
panel of the steam condenser on the back wall of the wash
cabinet because of angled wash items or injector nozzles.
When installing angled injector nozzles and sorting wash
items, ensure that the wash water outlets point towards
the center of the wash cabinet.
Restart the machine. Excess water is pumped out auto-
matically.
There is a problem with the process documentation inter-
face. The machine has detected a module for and Ethernet
interface, but only a serial interface is activated in the con-
trols (RS232).
Deactivate the RS232 interface:
Open the menu for configuring the interface
settings
Interface
/
Wait approx. 90 seconds. The Ethernet module XKM
3000 L Med needs this time for initialization. It may be
necessary to reconfigure the interface.
Or
Replace the Ethernet module XKM 3000 L Med with a
XKM RS232 10 Med module to set up a serial interface.
The peak-load cut-out has lasted longer than 3 hours.
Have your electrical system and your energy manage-
ment system tested by a suitably qualified person.
Frequently asked questions