Maintenance measures
Test programs
108
Various programs are available for monitoring the cleaning perfor-
mance in the course of routine checks. The test programs are not
separate processing programs. Rather, they are additional functions
that can be activated prior to starting any processing program.
The test programs interrupt the program sequence automatically at
specified points. The interruption is indicated by an audible signal
tone and message on the display. Miele Service can set the duration
of the interruption to between 10 seconds and approx. 42 minutes.
During this time period, measurements can be made or the door can
be opened to obtain a sample.
To prevent cooling of the wash cabinet, do not keep the door open
too long.
After the time period elapses, the program sequence continues auto-
matically. If the door has been opened, the program cannot start re-
sume until the door has been closed again.
If a measurement or sample is not needed, you can resume the pro-
gram sooner by pressing the Start/Stop button.
The following test programs can be selected:
Laboratory
-
The program sequence can be paused in each wash block immedi-
ately before the wash fluid is drained away.
Validation
-
The program sequence is interrupted at the following points:
– before the chamber washer solution is drained away in the final
wash block,
– after the interim rinse before the chamber washer solution is
drained away, and
– after water intake and before draining in the final rinse block.