Chapter 2: System Description

Data Outputs

The Analyzer has one Ethernet (TCP/IP and Modbus
ports in text, PDF, and CSV formats, allowing flexibility for exporting and printing data. Historical and diagnostic
data can be transferred from the Analyzer via the USB ports to a USB flash storage device (provided) and then
transferred to any computer that supports USB.
Real-time data can be transferred from the Analyzer via the Ethernet port to a computer using communication
software such as PuTTY
Data into a Spreadsheet."
Modbus are also provided in
outputs.
N
:
The Analyzer MUST be connected to a computer using an Ethernet cross-over cable or to a computer
OTE
network using an Ethernet straight-though cable. The Analyzer cannot be directly connected to a
computer via the USB port.

Sample Preparation Device Options

The Analyzer may be configured at the time of purchase to maximize performance by selecting a sample
preparation device most appropriate for the sample to be processed. The following table lists available sampler
configurations.
Clean Water Sampler
Wastewater Sampler
High Temperature
Sampler

Clean Water Sample Options

The Clean Water sample options include the 2-Stream Standard (standard or high-temperature option)
configuration and the 5-Stream (standard) configuration. The standard configurations accommodate sample
3.
Three (3) USB ports are provided for your convenience. One is located on the left side of the Analyzer and
the other two are on the System/Controller board. You can only export data from one port at any time.
The Analyzer will export to the first inserted USB flash storage device. While this USB flash storage device
is inserted, the Analyzer will not recognize any additional USB storage devices inserted in other USB ports.
GE Analytical Instruments ©2016
©
®
, HyperTerminal
, and Tera Term
Data also can be transferred through the Ethernet port using Modbus. Details on using
"Appendix D: Modbus Map."
Table 2: Sample Preparation Options
2-Stream
Standard
Optional
(Substitute for Clean Water Sampler)
Optional
®
3
) and three
USB ports. Data are output through the USB
©
. For details, see
Stream-specific data may be output via 4-20 mA
Standard
Optional
(Substitute for Clean Water Sampler)
Not Available
61 of 270
"Appendix C: Importing Analyzer
5-Stream
DLM 68100-09 EN Rev. A