Sony Trimaster PVM-L2300 Operation Manual page 45

Lcd video monitor
Hide thumbs Also See for Trimaster PVM-L2300:
Table of Contents

Advertisement

Menu
Contrast Hold
Auto
Auto Adjust
Target Color Temp
x
y
Target Luminance
Function and operation ([
Sets whether or not to adjust the color temperature and contrast (backlight) at the same time.
Off: Contrast is adjusted.
[On]: Contrast is not adjusted.
Adjusts the color temperature automatically by using an optional probe.
When the item is selected, the color temperature data to be adjusted is displayed.
When D93 or D65 data is changed, they are displayed as D93* or D65*.
When Contrast Hold is set to Off, the color temperature data and contrast data are displayed.
The contrast data is displayed as the picture preset data Preset1 to Preset5 set in Picture
Preset of the Input Configuration menu, or MANUAL. To change the displayed data from the
picture preset data to MANUAL, set the MANUAL button of CONTRAST to on. To change the
displayed data from MANUAL to the picture preset data, set the MANUAL button of
CONTRAST to off. To change the type of the picture preset data, select from Preset1 to
Preset5 in the Picture Preset menu of the Input Configuration menu.
For information about the Input Configuration menu, see page 48.
Starts adjustment.
Note
This menu is not selectable in the following case:
• When DVI Computer or the computer signal of HDMI is selected in the Format menu of the
Input Configuration menu
Before automatic adjustment, select the value to use for the adjustment in following
sequence.
When you use the previously adjusted values for adjustment, select Auto Adjust to start
adjustment without performing step 1, step 2 and step 3.
1 Select the color temperature to be used in Target Color Temp.
When Any is selected, enter the values of the CIE color system x and y coordinates.
2 When Contrast Hold is set to Off, select Target Luminance and enter the value of the
2
luminance (cd/m
) when the 100% white signal is input.
3 Select the probe to be used in Probe.
4 Select Auto Adjust.
The following message appears.
Set probe on cursor.
Attach the probe on the center of the display to start automatic adjustment. (It may take
some time until the automatic adjustment is completed.)
Note
First select the probe and then select Auto Adjust.
To cancel/stop the adjustment
Press the MENU button.
Sets the adjustment value to be used from D93, D65 or Any.
Enters the CIE color system x coordinate.
Set to 0.280 to 0.345.
Enters the CIE color system y coordinate.
Set to 0.280 to 0.352.
2
Enters the luminance (cd/m
) when the 100% white signal is input.
Adjust the luminance by the contrast adjustment (backlight adjustment).
Set to 40 to 250. [100]
Notes
• This menu is not selectable when Contrast Hold is set to On.
• The luminance is reduced to about half due to the principle of operation in the black frame
insertion mode or the interlace display mode. However, on this monitor the normal
luminance is set by the contrast adjustment (backlight adjustment). For this reason, as the
upper limit of Target Luminance becomes about 120 [cd/m
• The luminance calibrated internally by the backlight does not guarantee an absolute value
and may be different from the luminance calibrated by the probe. When you set Target
Luminance near to the upper limit (250) or the lower limit (40), the setting value may
exceed the adjustment range and cause an error. If an adjustment error occurs, change
Target Luminance and perform the automatic adjustment again.
]: factory setting)
2
], set it to less than 120 [cd/m
2
].
45