HP 37718A Installation And Verification Manual page 47

Communications performance analyzer
Hide thumbs Also See for 37718A:
Table of Contents

Advertisement

General Information
Specifications
POH: J1, C2, G1, F2, H4, Z3, Z4, Z5.
Any of the overhead bytes listed above
is selected and a BER measurement is
9
performed using a 2
-1 PRBS inserted
into a 64 kb/s channel. Single errors can
be added to the test pattern. Error
count, error ratio, error free seconds,%
error free seconds and pattern loss
seconds are measured.
Frequency measurement: Frequency
displayed in Hz, 1 Hz resolution. Offset
displayed in ppm and Hz.
Graphics and Logging
Graphic display or printout:
Barchart (results versus 60 time
periods) for current or stored
measurement period.
Bar resolution: 1 second, 1, 15, 60
minutes.
DSn payload bar graphs: Bit error
count, code error count, 140M frame
error count, 34M frame error count, 2M
frame error count, CRC error count,
REBE error count, DS3 frame error
count, DS1 frame error count, P-Bit
error count, C-Bit error count, DS3
FEBE, M2120 Reports and PDH
alarms.
SONET bar graphs: Frame error
count, CV-S error count, CV-L error
count, CV-LFE (REI-L) error count,
CV-P error count, CV-LFE (REI-P)
error count CV-IEC error count, CV-V
error count, CV-VFE (REI-V) error
count and SONET alarms.
STM-0/STM-1/STM-4/STM-16/OC-
1/OC-3/OC-12/OC-48 OPTICAL
INTERFACE (Option 104 - 1310nm.
Option 105 - 1550 nm. Option 106
1310 and 1550 nm).
Rate: 51.84 Mb/s, 155.52 Mb/s, 622.08
Mb/s, 2.48832 GHz.
Line code: NRZ
Connectors: FC/PC.
Transmit:
Wavelength: 1280 - 1330 nm and 1530
- 1570 nm.
Spectral width (3dB): <=0.3 nm rms
nominal (1310 and 1550 nm)
Spectral width (20dB): <=1 nm rms
nominal (1310 and 1550 nm)
Extinction ratio: >10 dB (1310 and
1550 nm).
Fibre power output: +1dBm typical
(1310 and 1550 nm). Minimum 0.0
dBm, Maximum +3.0dBm
Safety classification: Class 1, FDA 21
CFR Chapter1, Subchapter J and
IEC825-1.
Optical rise and fall times: Meets
ITU-T G.957 and Bellcore GR-253-
CORE.
Pulse mask: Meets ITU-T G.957 and
Bellcore GR-253-CORE masks at the
four standard rates when observed over
a 20s gating interval.
Receive:
Wavelength: 1280 - 1335 nm and 1500
- 1580 nm
Connector: FC/PC
1-37

Advertisement

Table of Contents
loading

This manual is also suitable for:

37719a