Equipment Required - HP 37718A Installation And Verification Manual

Communications performance analyzer
Hide thumbs Also See for 37718A:
Table of Contents

Advertisement

WA R N I N G
Performance Tests
Multirate Optical Interfaces
Description
The optical power output is measured on a power meter.
The receiver sensitivity is verified by attenuating the transmitter output
and checking for no errors in back-to-back mode.
The Protected Monitor Point (PMP) functionality is verified by looping
the HP37718A/19A Optical Output to the PMP input via an Optical
Attenuator and Lightwave Converter, then checking for error-free
operation.

Equipment Required

Power Meter
Power Meter Sensor
Module
Oscilloscope
Lightwave Converter
Optical Attenuator
FC/PC Connector Interface
Optical Cables (qty 2)
Adaptor (SMA to BNC)
Safety precautions must be observed when handling the
HP37718A/19A Optical Modules as these generate laser signals
which can cause serious injury. The guidelines below must be
followed:
Check the connector configuration of the Fiber Optic Interfaces.
If these are fitted with a connector interface other than FC/PC
then remove the existing connector interface and fit the FC/PC
connector interface.
Check for any damage to the HP37718A/19A Fiber Optic Interface
spring loaded aperture covers and connectors. Do not power up
the instrument if in any doubt about the integrity of these
connectors.
Make all connections to the HP37718A/19A Fiber Optic Interfaces
before powering up the instrument.
3-72
: HP8153A
: HP 81536A
: HP 54520A
: HP 11982A
: HP 8157A
: HP 81000FI (Qty. 4)
: HP 11871A
: HP 1250-1787