Performance Tests

SDH/SONET Transmitter Clock Accuracy

SDH/SONET Transmitter Clock Accuracy
Specification

Description

The test uses a Frequency Counter connected to the Multirate Analyzer
module Clock Trigger output port. This output is derived from the Clock
module internal 10MHz clock oscillator providing a 51.84 MHz that is
directly related to all SDH/SONET output rates.

Equipment Required

Frequency Counter

Procedure

1 Recall the HP37718A/19A DEFAULT SETTINGS as shown on Page 3-
2.
2 Set the Transmitter Output to SDH/SONET.
3 Connect the HP37718A/19A Multirate Analyzer module CLOCK
TRIGGER port to the Frequency Counter Input A, set the input
termination to 50
4 Adjust the Frequency Counter Trigger Level to obtain a stable reading
and ensure that the Frequency Counter reads between 51.839670
MHz and 51.840230 MHz.
5 Disconnect all the test equipment.
Bit Rate
Accuracy
51.84 Mb/s
4.5 ppm
: HP 5335A Option 010
.
3-51