Period (Per); Delay (Del) (After Trigger Out ); Double Pulse {Dbl) {Interval Between Leading Edges); Pulse Width {Wid) - HP 8112A Operating, Programming And Servicing Manual

50 mhz programmable pulse generator
Table of Contents

Advertisement

Period (PER)

Delay {DEL) {after
Tr igg e r Ou t)
Double Pulse {DBL)
{interval between
leading edges)

Pulse Width {WID)

Duty Cycle (DTY)

Linear Transitions
10%
{between
and
amplitude)
Note
2-2 Specifications
Range
20.0 ns to 950 ms
Ra ng e
75.0 ns to 950 ms (max:PER+55 ns)
Accuracy
±5% of programmed value ±5 ns
Range
20.0 ns to 950 ms (max:PER-WID)
DEL and DBL are mutually exclusive
Range
10.0 ns to 950 ms (max: PER- 10 ns)
Range
1 % to 99%, subject to Width specification
Resolution
1
Accuracy
±10% of programmed number
WID and DTY are mutually exclusive
90%
Range
6.5 ns to 95 ms, leading edge (LEE) and trailing edge (TRE)
independantly programmable within 1 :20 ratio
Delay, Width, and Transitions are under-programmable to ensure
that the specifi e d minimum values can always be obtained.

Hide quick links:

Advertisement

Table of Contents
loading