Power-down Detection
U2A is used to detect the power being switched off. Normally
its output is negative because its inverting input is at a higher
potential (5.1 V) than its non-inverting input (5 V). When the
HP 8 1 1 2A is switched off, the 5 . 1 V supply breaks down faster
than the 5 V reference because it is loaded by the microprocessor
board. This is detected by U2A which switches its output towards
its positive supply. This "Power Down Detected" signal is used
on the microprocessor board to ensure that the microprocessor
and HP-IB switch off cleanly. Refer to Chapter 10.7
Microprocessor.
Servicing the