HP 8112A Operating, Programming And Servicing Manual page 149

50 mhz programmable pulse generator
Table of Contents

Advertisement

PERFORMANCE TEST RECORD: Hewlett-Packard 8 1 1 2A 50 MHz Pulse Generator
Serial No:
- - - - - - - - - - - - - - - - - - - - - - - - - - -
Transition Times
Transition
Period Leading Edge Trailing Edge Minimum MEASURED Maximum Uncertainty
Fixed
100 ns 4.5 ns
Linear
100 ns 6.5 ns
Period Leading Edge Trailing Edge Minimum MEASURED Maximum Uncertainty
2 µs
500 µs 100 µs
2 ms
5 ms
5 ms
50 ms
Report No: -------------------------- Date: -------------
Low
-0. l v
Low-Level Minimum MEASURE.D Maximum Uncertainty
-0.5 v
- 1 .0 v
-5.0 v
-8.0 v
Fast
4.5 ns
6.5 ns
Slow
500 ns
500 ns
100 µs
500 µs
500 µs
999 µs
999 µs
1 ms
1 ms
10 ms
lO ms
PERFORMANCE TEST RECORD Page 5 of 6
-0.056 v
-0.44 v
-0.92 v
-4.76 v
-7.64 v
-
4.675 ns
473 ns
95 µs
475 µs
949 µs
0.95 ms
9.5 ms
-0. 144 v
-0.56 v
- 1 .08 v
-5.24 v
-8.36 v
(typ)
5.0 ns
8 .825 ns
527 ns
105 µs
525 µs
l.049ms
1 .05 ms
10.5 ms
Testing Performance 8-37