5. Record your results on a i:opy of the Test Record, specified limits
are given here and on the Test Record.
The oscilloscope trace flatness error may affect the measurement of
Note
pre- and overshoot.
6. Change HP 8 1 12A settings as follows:
Transition
LEE
TRE
Examine the HP 8 1 12A output on the scope display in order to
7.
verify that the pulse characteristics do not exceed the specified
limits as identified in Figure 8- 1 1.
8 . Record your results on a copy of the Test Record, specified limits
are given here and on the Test Record.
The oscilloscope trace flatness error may affect the measurement of
Note
pre- and overshoot .
Characteristic
Preshoot
::;±10% of amplitude ±10 mV
Overshoot/Ringing ::;±10% of amplitude ±10 mV
Settling time
::; 105 ns
Linear
6.5 ns
6.5 ns
Characteristic
Preshoot
::;±5% of amplitude ±10 mV
Overshoot /Ringing ::;±5% of amplitude ±10 mV
Settling time
::; 107 ns
Specification
Specification
Testing Performance
8-17