Apexpro Fh Telemetry System Component Configuration; Process Overview; Set The Apexpro Fh Telemetry System Settings; Complete Installation Checkout Procedures On Page - GE ApexPro Service Manual

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Configuration
Complete installation checkout procedures
ApexPro FH Telemetry System component
configuration

Process overview

Set the ApexPro FH Telemetry System settings

5-18
Proceed to
Installation checkout procedures on page
After completing the physical installation of the equipment and you are able to access
the ATS, you can configure the standard and specialized functions of the ApexPro FH
Telemetry System components. Perform the following procedures:
Access the ATS on page
Configure the laptop on page
Connect the service laptop to the Unity Network IX network on page
Set the ApexPro FH Telemetry System settings on page
Set the ApexPro FH Telemetry System network name and IP address on page 5-
20.
Verify the IP address settings (if required) on page
Enable access points for activity on the network on page
Configure the server for Remote Service access on page
Configure the Transmitter battery status report for viewing and printing on page
5-13
Back up the configuration settings on page
Perform installation check on page
Complete
Installation checkout procedures on page
Perform this procedure to set the following settings:
Language
Mode
Tower name
Unit name, if required
A/C power frequency
Device Master designation
(Optional) Print summary report when patient is discharged
Localization, if required
1.
Log onto the Webmin service interface on page
2.
Select the Configuration tab.
ApexPro™ Telemetry
8-2.
5-2.
4-3.
5-5.
5-15.
5-16.
4-7.
B-4.
5-18.
5-22.
5-10.
8-2.
2001989-203C