Install The Software On The Target Atss - GE ApexPro Service Manual

Telemetry server
Hide thumbs Also See for ApexPro:
Table of Contents

Advertisement

Install the software on the target ATSs

2001989-203C
c.
Select File > Save. Specify the destination and file name of this .txt file so
you can easily navigate to it.
d.
From the GE Healthcare Software Transfer Utility window, select File >
Import IP File. Navigate to the .txt file containing the IP addresses. The IP
addresses should display in the Add Target IP Addresses list.
3.
Verify these are the ATSs you want updated.
NOTE
All IP addresses displayed in the GE Healthcare Software Transfer Utility
> Add Target IP Address list will have the selected software packages
installed on them. Verify the list of target IP addresses only contains the
ATSs you want updated.
4.
To remove an IP address from the displayed list, select the IP address and click
(up arrow).
Once the GE Healthcare Software Transfer Utility > Add Target IP Address list only
displays the IP addresses of the ATSs you want to update, you are ready to start
installing the software packages.
Complete the following procedure to install your selected software packages on the
target ATSs:
1.
Under Select Software Packages, choose the software packages you want to
install on the target ATSs.
NOTE
Ctrl + click to select multiple software packages.
2.
Select Transfer. Transfer status information displays in the Transfer Status
window at the bottom of the GE Healthcare Software Transfer Utility window.
ApexPro™ Telemetry
Upgrading ATS software
657A
B-7