HP 5890 SERIES II Service Manual page 26

Gas chromatograph
Hide thumbs Also See for 5890 SERIES II:
Table of Contents

Advertisement

SEPTUM- PURGED PACKED COLUMN INLET
Remove/Replace Septum- Purged Packed Column Inlet
HAZARDOUS VOLTAGES ARE PRESENT IN
THE INSTRUMENT WHEN THE POWER CORD
IS CONNECTED.
DANGEROUS
DISCONNECTING
BEFORE WORKING ON THE INSTRUMENT.
1. Set the main power line switch to the off position.
2. Disconnect the power cable from its receptacle.
3. Allow time for the oven and heated zones to cool.
4. When the heated zones are cool, turn off all gas
supplies.
5. At the bottom of the inlet(s) to be removed, in-
side the column oven, remove the column and
hardware associated with the inlet(s) (liner,
column/liner nuts, ferrules, makeup gas adapt-
er, etc.).
6. Remove the injection port cover by grasping its
back edge and lifting it upward.
7. Remove the two screws securing the left side
panel along its bottom edge.
8. Slide the left side panel towards the
rear of the instrument and lift.
SCREWS
9. Disconnect the Carrier Gas Inlet (labelled " C" )
which connects to a fitting either at the EFS, or
at a mass flow controller mounted on the flow
panel.
10. Disconnect the Septum Purge Outlet (labelled
" P" ) which connects to a pressure regulator
(" IN"fitting) mounted on the flow panel.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
AVOID A POTENTIALLY
SHOCK
HAZARD
THE
POWER
CORD
LID SHAFT
LEFT
SIDE
PANEL
BY
INJECTION PORT COVER
INJECTION
PORT
PURGE TUBE
SVC 2- 5
RED O VISIBLE
OFF
ON
TOP COVER
PRESS
HERE TO
FREE
COVER
1/4- INCH
SCREW
AND
WASHER
INJECTION
PORT
CARRIER
TUBE