HP 5890 SERIES II Service Manual page 141

Gas chromatograph
Hide thumbs Also See for 5890 SERIES II:
Table of Contents

Advertisement

11. Release the extension spring securing the PMT assembly to its support bracket.
12. Remove the photomultiplier tube (PMT) assembly and sulphur filter from the detector assembly
and set it aside.
(CHIMNEY AND BRACKETRY
REMOVED FOR CLARITY)
13. Remove the drip tube from the exhaust tube at the top of the detector.
14. Remove the exhaust tube from the detector using a
9/16- inch wrench.
15. Remove the chimney assembly by removing
the two screws securing it to the chimney back.
16. Loosen the three screws which secure the
clamp which secure the detector to the chim-
ney back.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
SULPHUR FILTER
PMT ASSEMBLY
CLAMP
SECURING
SCREWS
SVC 4- 71
CHIMNEY
SCREW