7. The result appears on the
display, along with the control
bottle symbol and flashing "L."
Do not remove the test strip
yet. Press
once to mark the
result as a Level 1. Press
second time to mark the result
as a Level 2.
(example)
8. Press and release
control level in the meter.
a
(example)
9."OK" and the control result
to set the
alternate on the display if the
result is in range. The range is
printed on the test strip
container label. "Err" and the
control result alternate on the
display if the result is out of
range. Remove and discard the
used test strip per facility
policy.