GE moisture.iq User Manual page 79

Multi-channel moisture and oxygen analyzer
Hide thumbs Also See for moisture.iq:
Table of Contents

Advertisement

5.2.2 Entering Calibration Data Manually (cont.)
2. Just below the probe serial number, tap the Channel button to select the channel to which the probe is
connected.
3. To the right of the Channel button, tap the Type button to select the type of input connected to that
channel ( Hygro, Pressure, Oxygen, Aux 1 or Aux 2 ).
4. For each input selected, the table below the two buttons includes data entry boxes for 2 to 16 data
points. The four buttons to the right of the table are used to edit and arrange the calibration data as
required:
Sort - Tap this button to sort the data rows in ascending order by the left column.
Edit Value - Tap this button and then tap a text box and enter the values from the relevant Calibration Data
Sheet for each point specified, using the calculator keys on the pop-up keypad. Continue entering or checking
values for each input until you have completed each channel.
Insert Row - Use this button along with the Remove Row button to organize the rows in the table as desired.
Remove Row - Use this button along with the Insert Row button to organize the rows in the table as desired.
5. After you have finished entering your probe calibration data, tap the Return button to return to the
Configuration Menu.
moisture.IQ User's Manual
Figure 58: Probe Calibration Screen
Chapter 5. Using the Configuration Menu
67