GE moisture.iq User Manual page 148

Multi-channel moisture and oxygen analyzer
Hide thumbs Also See for moisture.iq:
Table of Contents

Advertisement

Appendix D. Modbus Register Map
Function Parameter
Channel 5 /
Alarm B
Channel 6 /
Alarm A
136
Table 13: Modbus Register Map (cont.)
Sub
Modbus
Parameter
ID
Address
Lower
508 3508
Setpoint
Upper
510 3510
Setpoint
Status
520 3520
State
522 3522
Units
524 3524
Type
526 3526
Lower
528 3528
Setpoint
Upper
530 3530
Setpoint
Status
600 3600
State
602 3602
Units
604 3604
Type
606 3606
Type
Access
32-bit float
RW
Read
Holding
(0x03),
Write
Multiple
(0x10)
32-bit float
RW
Read
Holding
(0x03),
Write
Multiple
(0x10)
32-bit Integer RO
Read Input
(0x04)
32-bit Integer RO
Read
Holding
(0x04)
32-bit Integer RW
Read
Holding
(0x03)
32-bit Integer RW
Read
Holding
(0x03)
32-bit float
RW
Read
Holding
(0x03),
Write
Multiple
(0x10)
32-bit float
RW
Read
Holding
(0x03),
Write
Multiple
(0x10)
32-bit Integer RO
Read Input
(0x04)
32-bit Integer RO
Read
Holding
(0x04)
32-bit Integer RW
Read
Holding
(0x03)
32-bit Integer RW
Read
Holding
(0x03)
Modbus
Qty. of
Request
Registers
2
2
2
1 = Energized
0 = De-energized
2
0 = Disabled,
1 = Enabled
2
Measurement/Unit
Code 0 = off,
others see tab
2
1 = Setpoint,
2 = In Band,
3 = Out Band
2
2
2
1 = Energized
0 = De-energized
2
0 = Disabled,
1 = Enabled
2
Measurement/Unit
Code 0 = off,
others see tab
2
1 = Setpoint,
2 = In Band,
3 = Out Band
moisture.IQ User's Manual
Notes