Chapter 1. Installation
1.9.4.2 Weatherproof Delta F Oxygen Cells
The weatherproof oxygen cell has a set of sensing and secondary electrodes that are wired to a terminal strip in
the weatherproof enclosure. Connect the weatherproof oxygen cell using a four-wire shielded cable with a
mating bayonet-type connector. Fasten the bayonet-type connector to the mating connector on the bottom of
the weatherproof enclosure. Connect the other end of the cable to the OXYGEN terminal block on the back of
the moisture.IQ. Refer to Figure 18 on page 22 and Table 3 below for the required connections.
green wire (–)
white wire (+)
black wire (–)
24
Table 3: Weatherproof Delta F Oxygen Cell Connections
To Delta F Enclosure
Connect:
Terminal Block
red wire (+)
To moisture.IQ OXYGEN
Terminal Block
pin 1
pin 2
pin 3
pin 4
pin 4
pin 3
pin 2
pin 1
moisture.IQ User's Manual