Appendix D. Modbus Register Map
Function Parameter
142
Table 13: Modbus Register Map (cont.)
Sub
Modbus
Parameter
ID
Address
Hygro - Equiv.
206 5306
Dew Point °C
Hygro - Equiv.
208 5308
Dew Point °F
Hygro - PPMv 210 5310
Hygro - PPBv 212 5312
Hygro - PPMw 214 5314
Hygro - RH % 216 5316
Hygro -
218 5318
Lbs/MMSCF
(ideal gas)
Hygro -
220 5320
Lbs/MMSCF
(natural gas)
Hygro - PPMv
222 5322
(natural gas)
Hygro - g/m^3 224 5324
Hygro -
226 5326
mg/m^3
Hygro - Vapor
228 5328
Pressure (kPa)
Hygro - Vapor
230 5330
Pressure
(mm Hg)
Hygro - MH /
232 5332
FH
Temperature -
234 5334
°C
Temperature -
236 5336
°F
Temperature -
238 5338
Kelvin
Temperature -
240 5340
Rankine
Pressure - kPa
242 5342
-gauge
Pressure -
244 5344
Mpa - gauge
Pressure - Pa
246 5346
-gauge
Pressure - Bar
248 5348
-gauge
Type
Access
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
32-bit float
RO
Read Input
(0x04)
Modbus
Qty. of
Request
Registers
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
moisture.IQ User's Manual