Weighing Functions

5.2.21 DR3 - Maximum weight for weighing range 3

The weight can be used for calibrating purposes with the numeral step for
measurement range 3 only under the maximum weight. The maximum weight is
defined with the adjustment.
The maximum weight depends on the number and type of load cells used.
Note
If only one or two ranges are defined then this entry is meaningless.
5.2.22 DR3 - Numeral step for weighing range 3
The numeral step for weighing range 1 can be defined according to standard
EN 45501 (0.0001 to 50)
Note
If only one or two ranges are defined then this entry is meaningless.
5.2.23 DR3 - Stand-still time 1
Stand-still monitoring is used for recognising a stable scale balance.
The scale stand-still is defined if the weight value is changed within a defined time
(stand-still time) by less than a defined deviation range (Stand-still value). Stand-
still 1 is used in static scale operation (Instructions: Zero setting, Taring, Log
output).
5.2.24 DR3 - Stand-still range 1
Stand-still monitoring is used for recognising a stable scale balance.
The scale stand-still is defined if the weight value is changed within a defined time
(stand-still time) by less than a defined deviation range (Stand-still value). Stand-still
1 is used in static scale operation (Instructions: Zero setting, Taring, Log output).
The functionality of the stand-still monitor clarifies the following image.
5-54
SIWAREX FTA