62
2.20
Checking the kV and mAs displays
•
Check whether the green "ready" lamp lights up approximately 2 minutes after switch-
ing the system on. If it doesn't light up, an error message will appear on the display. Re-
fer to the chapter "Display messages" in the Service Instructions.
•
Check whether the kV and mAs values are clearly visible from the operator's position.
Determine whether all exposure parameters are correctly displayed.
PMF
"Ready" indicator
SIE
Visibility of displays
SIE
Correct displays
MOBILETT Plus HP
SPR8-220.101.04.03.02
07.05
Inspection and Maintenance
Page 62 of 64
CS PS 24
Siemens AG