2.16

Checking the manual exposure interrupt

The user must be able to interrupt the exposure at any time.
Set 70 kV, 160 mAs.
X
Release an exposure and immediately interrupt it by letting go of the exposure switch.
The message "USE 20" appears on the display indicating that the exposure release
switch was interrupted during the exposure. In addition, 10 short "beep" signals sound.
The message can be erased by pressing the "kV+" button.
SIE
Message "USE20"
SIE
10 short beep signals
NOTE
MOBILETT Plus HP
All covers must be remounted.